Home ] Contact Us ] Search ]  

 

 

 

 

Home
About
News
Products
Support
Feedback
Career

 

 

SARS UPDATE

SARS UPDATE

 
Download PDF Ver.

In response to the many inquires we’ve received regarding the use of filters with Impact products, we can provide you with the following information:

  1. Many filter products are available and most are capable of attachment to the Impact family of aspirator and ventilator products.  To date, none of these filters have been represented by their manufacturers as being effective against the SARS virus.  Accordingly, Impact will make no specific recommendation regarding their efficacy beyond those representations already made by their respective manufacturers.

  1. Our engineering staff has tested the filter types listed in the tables below.  This allows us to determine their effect upon the performance of our products and in turn gives our users necessary information should the filter’s effect upon product performance be of concern to them.

  1. On behalf of our customers, Impact will accept filter products for test as was done below.  We will test these filters only if they’re provided in unused condition, clearly labeled/marked, and contained within their original packaging.  The results of our tests will be added to the tables below.

  1. Given the uncertainty of stocking requirements and customer need, Impact will not inventory these filters.  Filters should be purchased from the respective manufacturer or through its authorized dealer(s). 

Filter Brand & Part Number

Impact Adapter Kit Part Number

Tested for use with the following Impact products

Effect upon performance

Pall 8001052

714-0013-10

Portable Aspirators:  305, 308, 1305, 320, 321, 326

On-board Aspirators:  303, 309, 319, 324

Reduces airflow by 7 to 8 liters per minute.  No effect on vacuum.

 

 

 

 

Pall 8001052

540-0029-35

Sorensen:  1120, 2110, 2121, 2135, 2590, 2591, 1666, 3666

Reduces airflow by 7 to 8 liters per minute.  No effect on vacuum.

 

 

 

 

Pall BB50T

540-0188-10

Sorensen:  2492

None

 

 

 

 

Pall BB50T

None Required

Uni-Vent:  750, 754

None

 

 

 

 

Pall InterveneTM 7 Filter

Portable Aspirator 326, 325, 321, 320

Reduces airflow by 9 to 10 liters per minute. No effect on vacuum
  1. Filter information may be obtained from the Pall website:  http://www.pall.com/catalogs/

Click on:  Medical Filtration Solutions

Click on:  Respiratory Care

Click on:  Conserve Breathing Circuit Filter (this is the BB50T filter)

Note:  the Pall 8001052 filter is new.  At the time of this posting it did not appear on their website.

ASPIRATOR ADAPTER KIT MATRIX

Impact Model/Series Number

Tubing (I.D.)

Adapter Kit

Part Number

Contents/Kit

 

 

 

 

305 Portable

3/8

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

308 Portable

3/8

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

1305 Portable

3/8

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

320 ultra-lite Portable

5/16

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

321 ultra-lite Portable

5/16

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

326 ultra-lite Portable

3/8

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

 

 

 

 

303 On-board

3/8

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

309 On-board

3/8

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

319 On-board

3/8

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

324 On-board

3/8

714-0013-10

20 ea. ผ” x 3/8” Adapters; 10 ea. 3/8” ID x 12” Tubing

 

 

 

 

Sorensen Suction

 

 

 

1120 General Purpose

1/4

540-0029-35

35 ea. ผ” ID x 12” Tubing

2110 General Purpose

1/4

540-0029-35

35 ea. ผ” ID x 12” Tubing

2120 General Purpose

1/4

540-0029-35

35 ea. ผ” ID x 12” Tubing

2135 General Purpose

1/4

540-0029-35

35 ea. ผ” ID x 12” Tubing

2590 Thermotic

1/4

540-0029-35

35 ea. ผ” ID x 12” Tubing

2591 Thermotic

1/4

540-0029-35

35 ea. ผ” ID x 12” Tubing

1666 Uterine

1/4

540-0029-35

35 ea. ผ” ID x 12” Tubing

3666 Uterine

1/4

540-0029-35

35 ea. ผ” ID x 12” Tubing

 

 

 

 

2492 Thoracic

3/4

540-0188-10

10 ea. พ” ID x 12” Tubing

VENTILATOR INFORMATION

Impact Model/Series Number

Exhalation Valve Port (O.D.)

Recommended

Filter

Notes

 

 

 

 

750 Uni-Vent

22mm

Pall BB50T

No adapter required, connect directly to disposable circuit’s exhalation valve port.

754 Uni-Vent Eagle

22mm

Pall BB50T

No adapter required, connect directly to disposable circuit’s exhalation valve port.


 
 
 

COPYRIGHT 2002 Impact Instrumentation, Inc. Privacy Policy